Correlations between the infrared-AI software test results and patient characteristics
Clinical variables | Value | Infrared-AI software result | Infrared-AI software result | χ2 | P value | ||||
---|---|---|---|---|---|---|---|---|---|
FP | FN | Total | TP | TN | Total | ||||
BMI (kg/m2, n = 349) | Underweight | 0 | 0 | 0 | 0 | 2 | 2 | 1.81 | 0.697 |
Normal | 16 | 0 | 16 | 7 | 44 | 51 | |||
Overweight | 37 | 1 | 38 | 15 | 109 | 124 | |||
Obesity | 33 | 1 | 34 | 15 | 69 | 84 | |||
Breast pain | No | 74 | 0 | 74 | 17 | 184 | 201 | 1.97 | 0.16 |
Yes | 12 | 2 | 14 | 20 | 40 | 60 | |||
Activities and conditions that affect body temperature | No | 27 | 0 | 27 | 17 | 48 | 65 | 1.13 | 0.287 |
Yes | 59 | 2 | 61 | 20 | 176 | 196 | |||
Breast density | A | 7 | 0 | 7 | 5 | 14 | 19 | 7.67 | 0.081 |
B | 65 | 2 | 67 | 21 | 168 | 189 | |||
C | 12 | 0 | 12 | 11 | 42 | 53 | |||
D | 2 | 0 | 2 | 0 | 0 | 0 | |||
BC grade (n = 39) | I | - | 0 | - | 1 | - | - | 0.39 | 0.811 |
II | 1 | 19 | |||||||
III | 1 | 13 | |||||||
N/A | 0 | 4 | |||||||
Tumor size | T1 | - | 0 | - | 19 | - | - | 6.02 | 0.120 |
T2 | 1 | 15 | |||||||
T3 | 1 | 2 | |||||||
T4 | 0 | 1 |
Error is the false-positive (FP) and false-negative (FN) given by the infrared-AI software; thus, hit is the true positive (TP) and true negative (TN). N/A: information not available; -: bank cell; *: P value calculated by Fisher’s exact test